3. Press“ ” to start infusion. The LCD“ ” moves indicating the infusion is running;
Attention: Observe if the drop detector lights for each drop detected
3.7 Finishing Infusion
1. When volume infused display reaches volume to be infused (limit) setting, the pump will display a
” alarm to notify that the infusion is completed. The pump will turn to KVO function. The
minimum flow rate is maintained to prevent occlusion. Press“ ” to stop KVO function.
3.8 Powering Off
1. Press“ ”for 0.5 seconds until powered off;
3.9 Extended Operation
3.9.1 Pause/Restart
a) Pause infusion
: pressing the “ during infusion will pause the pump.
b) Restart:
pressing the “ ”again, the pump will restart the infusion.
NOTES: Pausing for more than 2 minutes will activate the alarm system (Pause Overtime)
Press “ to clear the alarm and maintain the pause state.
3.9.2 To continue another solution container
1. Change to a new infusion container and make sure that the IV set is connected correctly.
2. Change the IV tubing down or up 20cm to make the linear pump cassette act on a new position
of tubing after the IV set has been used over 5-6 hours.
3. The total infused volume will add with the new volume to be infused programmed. If you want
to clear the total infused volume display:
With the “ ”flashing, press“ ” for 0.5 seconds and the display will reset to 0.