EasyManua.ls Logo

DLab D2012 - Operation; Normal Operation; Rotor Loading and Removal

DLab D2012
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL
10
8. Operation
CAUTION
Do not push or lean against the centrifuge while it is running.
Do not run the centrifuge when fragments or sample solutions are left in the centrifuge chamber.
Always keep the centrifugal chamber clean.
If the centrifuge makes strange noise during operation, stop it immediately and contact our service
center. Notify them of the warning code if displayed.
8.1 Normal Operation
Turn on the power switch, centrifuge will display the running interface last time after passing the
self-diagnostic checks, see figure 8-1 below:
Figure 8-1 the last running interface
Speed: 15000rpm. Running time: 12 minutes.
The door lock is released.
8.1.1 Rotor loading and removal
CAUTION
Attach the rotor to the rotor shaft. Ensure the rotor is in position and connected with the shaft,
tightening the locking nut to secure the rotor with shaft, to prevent the rotor damaging the centrifuge.
Ensure the rotor lid is firmly tightened to the rotor.
150 12 min
×100rpm

Related product manuals