11.1.3 Notice
The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated samples
density is over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
11.1.4 Autoclaving
A12-2 rotor is manufactured in high-strength aluminum alloy material and can be autoclaved: 121 (1.0kg/cm2),
20 minutes.
A12-2P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.
CAUTION:
The lid of the rotor is made of plastics, can not be high-pressure sterilization, only ordinary sterilization
can be used.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
Table 11.2 Cleaning and sterilizing conditions for tubes
OApplicable XInapplicable
Condition Material
PA
PC
PP
Cleaning
Cleaning fluids
AcidicpH5 or lower
X
X
X
Acidichigher than pH5
O
O
O
Alkalinehigher than pH9
O
X
O
AlkalinepH9 or lower
O
O
O
NeutralpH7
O
O
O
Warm water(up to 70)
O
O
O
Ultrasonic cleaning
Neutral detergentpH7
O
O
O