Table 10-1 Frequent problems and solutions
Warning code E-1~9 are related to wrong operating. You can continue running the centrifuge after the
cause removed.
10.2 How to open the door
10.2.1 In the case of power on
(1)Turn on the POWER switch, the door lock will release automatically.
(2)The door lock will release automatically once the operation finished.
(3)It is available to release the door by pressing button once the rotor stops.
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually.
(1)Ensure if the rotor has stopped rotating.
Listen carefully to ensure no rotating sound can be heard.
(2)Insert a screw driver into the hole to open the door.
Hole is located on the top right side of the unit.
Insert a screw driver into the hole and push forward to release the door.