High Speed Mini Micro-Centrifuge
15
·Rotor lid loosed
·Tighten the rotor lid firmly
Alarm code appeared on the time
display screen
E-02
Door fault
·The door opened in running.
· The button is pressed
while the door opening.
·Close the door immediately.
·Close the doorand then start
to operate.
E-06
Set wrong speed
·The setting speed exceed the
allowable range.
·Modify the speed value.
E-1086
·Read the service manual
·Contact with service center
Table 10-1 Frequent problems and solutions
Warning code E-19 are related to wrong operating. You can continue running the centrifuge after the
cause removed.
10.2 How to open the door
10.2.1 In the case of power on
1Turn on the POWER switch, the door lock will release automatically.
2The door lock will release automatically once the operation finished.
3It is available to release the door by pressing button once the rotor stops.
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually.
1Ensure if the rotor has stopped rotating.
Listen carefully to ensure no rotating sound can be heard.
2Insert a screw driver into the hole to open the door.
Hole is located on the top right side of the unit.
Insert a screw driver into the hole and push forward to release the door.
CAUTION
The door just can be opened while the power on and rotor stops rotating.