13
6. Function: Heating (LCD digital hotplate model)
The device is controlled by digital temperature control technology, which has two
separate safe circuits. The hotplate is kept at a constant temperature by a digital
control circuit. The hotplate temperature can also be monitored from a separate,
adjustable safe circuit. The two temperature sensors (PT 1000) internal for
temperature control are built into the hotplate. The single external PT 1000 can
monitor the temperature of sample.
Plug in the external PT1000.
Set the temperature via rotating the temperature control knob slowly to the
target value.
When the heating function is switched on, the LED “Heat” will light up and the
LCD will display the real temperature.
The rated temperature will be displayed on the right-hand side of the LCD as
well as Temp and °C characters.
The heating function is switched on or off by pushing heating knob.
Instrument is turned on for the setting zone temperature setting when the last
shutdown temperature, the common case, set the heating temperature display the
actual temperature may have the following differences: