18
Describe& conditions code conditions
The temperature sensor is
incorrectly placed or not put
into the container during
heating.
The temperature rises less
than 2 °C
heating
ER9
1. he setting temperature shall be
more than the initial temperature
displayed on the screen +10℃
2. The initial temperature displayed
on the screen shall be less than 35℃.
3. The temperature shall rise within
2℃ after heating for 7min.
If these faults are not resolved, please contact dealer/ suppler.