Operation
Using
skin
temperature
control
Connecting
skin
temperature
sensor
1
Insert
yellow
sensor
plug
into
the
yellow
socket
on
left
side
of
housing.
2
Route
sensor
cable
through
one
of
the
flexible
hose
seals
in
the
canopy.
3
Remove
protective
foil
from
adhesive
pad
and
place
skin
temperature
sensor
on
pad.
Using
the
adhesive
pad,
attach
sensor
tip
to
the
infant's
skin
where
the
temperature
is
to
be
measured.
Keep
sensor
cable
in
place
with
adhesive
plaster.
Placing
sensor
If
the
baby
is
lying
on
its
back:
Attach
sensor
to
the
abdomen
in
the
region
of
the
liver.
If
the
baby
is
lying
on
its
stomach:
Attach
the
sensor
to
the
back
in
the
region
of
the
kidneys.
The
sensor
must
never
be
attached
under
the
baby
since
it
would
then
be
measuring
and
controlling
core
temperature
and
not
skin
temperature.
Check
regularly
that
the
skin
temperature
sensor
is
properly
fixed
to
the
baby's
skin.
A
skin
temperature
sensor
which
has
fallen
off
would
be
measuring
air
temperature
so
that
the
baby
could
become
overheated
(though
the
temperature
of
the
air
in
the
incubator
would
not
exceed
39
°C).
Do
not
use
skin
temperature
sensors
to
measure
rectal
temperature.
When
a
skin
temperature
sensor
is
attached
but
the
"air
temperature
control"
mode
is
in
operation,
the
measured
skin
temperature
is
displayed.
Skin
temperature
is
not
then
being
controlled.