EasyManua.ls Logo

Dräger Aerotest HP - Page 30

Dräger Aerotest HP
148 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
Measurements in N
2
O
Pressure: 3 bars ±0.2
Measurements in CO
2
Pressure: 3 bars ±0.2
Measuring point
in the
measuring device
for
Tube
Limit
value
Flow
(l/min)
±10 %
Test time
(min)
Evaluation
Display Measurement range
Deviation from
standard
CO
2
CO
2
100/a-P 300 ppm 0.167 6 ppm 100 to 3000 ppm ±10 to 15 %
NO
x
NO
x
0.2/a 2 ppm 0.167 3 ppm NO
x
0.2 to 6 ppm ±30 %
CO CO 5/a-P 5 ppm 0.167 6 ppm 5 to 150 ppm ±10 to 15 %
H
2
OH
2
O 20/a-P 60 ppm
(45 mg/m
3
)
3.33 12 mg/m
3
20 to 250 mg/m
3
±10 to 15 %
Measuring point
in the
measuring
device for
Tube Limit value
Flow
(l/min)
±10 %
Test time
(min)
Evaluation
Display Measurement range
Deviation from
standard
SO
2
SO
2
1/a 2 ppm 0.167 12 Evaluation on scale of n = 10
Display x 0.2
= ppm SO
2
0.5 to 2 ppm ±30 %
H
2
SH
2
S 1/d 1 ppm 0.167 6 Evaluation on scale of n = 10
ppm H
2
S 1 to 20 ppm ±15 %
NO
x
NO
x
0.2/a 2 ppm 0.167 3 ppm NOx 0.2 to 6 ppm ±30 %
CO CO 5/a-P 5 ppm 0.167 6 ppm 5 to 150 ppm ±10 to 15 %
H
2
OH
2
O 20/a-P 60 ppm
(45 mg/m
3
)
3.33 12 mg/m
3
20 to 250 mg/m
3
±10 to 15 %

Table of Contents