9.3 Attachment 3: Dräger Fabius Family with Installed Fresh Gas Outlet (FGO)
FGO variant A FGO variant B
Different variants of the FGO exist. Please ensure that the switch of FGO variant A is at the COSY
position, as shown in the picture. We also recommend that you secure the switch in the COSY position
to prevent unintentional switching, by using tape for example. In Variant B, please ensure that the fresh
gas connection from the FGO to the COSY breathing system is installed properly.
For further details, please refer to the IFU and watch: https://www.youtube.com/watch?v=Sw5idUnpIZg