(a) If a shortage of the fresh gas delivery occurs (fresh gas flow < 150% of minute volume)
rebreathing will increase with a significant inspiratory CO2 and reduced FiO2 concentration.
(b) FiO2 will decrease due to entrainment of ambient air. No manual support of patient breathing
via Man / Spon possible. No volatile agents must be used.
(c) FiO2 will drop immediately. FiO2 may fluctuate. No manual support of patient breathing via Man
/ Spon possible. No volatile agents must be used. Take care that the connector for the absorber
is not blocked (by e.g. a plastic bag).
Setup when CO2-absorber incl. Clic-adapter is removed:
Apollo Fam.
Fabius Fam.
Perseus
7 Only soda lime of lower quality (granular soda lime) is available
Use a dust filter on the reusable absorber
Part number: MX50115 (or as an alternative bacterial filter IBF MK02588)
Exchange the filter with every change of the soda lime
8 No fresh water trap for the gas measurement (Waterlock2) is available
Empty the water trap regularly, at the latest when it is half full
In this case, continue using the water trap as long as possible
Even if the water trap is blocked (swelling elements are soaked), leave it in place to protect the
gas measuring bench. Due to the blocked water trap, there is no gas measurement available.
Therefore, the fresh gas flow must be set to at least 150% of the MV. Volatile agents must not
be used. Check alternative monitoring solutions (e.g. frequent blood gas analyses, narrow
alarm limits for hemodynamic monitoring).

Table of Contents

Other manuals for Dräger Fabius Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Series and is the answer not in the manual?

Dräger Fabius Series Specifications

General IconGeneral
TypeAnesthesia Workstation
ManufacturerDräger
Ventilation ModesSIMV
Flow ControlElectronic
Tidal Volume Range20 to 1400 mL (model dependent)
Respiratory Rate4 to 60 breaths/min
PEEP0-20 cmH2O
Gas SupplyO2, Air, N2O
Vaporizer Mounts2
MonitoringO2, CO2, N2O
DimensionsVaries by model
WeightVaries by model
ModelFabius Series
CategoryAnesthesia

Related product manuals