9 No sample gas line is available
Use infusion line or CO
2
measurement line with LuerLock
Gas measurement might be affected, especialy when applying higher respiratory rates and
lower tidal volumes
Alarms like „sample line blocked“ or „sample line disconnected?” might appear
10 No breathing system filter is available
General information on the filter setup and necessary reprocessing measures are in the customer letter
“SARS-CoV-2 and handling of Dräger Anesthesia Workstations“, https://www.draeger.com/en-
us_us/Home/novel-coronavirus-outbreak
Overview of Dräger filter / HME types and abbreviations for the decision tree:
MF/HME Mechanical filter with HME = Dräger TwinStar HEPA
MF Mechanical filter = Dräger SafeStar
EF/HME Electrostatic filter with HME = Dräger TwinStar
EF Electrostatic filter = Dräger CareStar
HME HME only (no filter) = Dräger HumidStar

Table of Contents

Other manuals for Dräger Fabius Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius Series and is the answer not in the manual?

Dräger Fabius Series Specifications

General IconGeneral
TypeAnesthesia Workstation
ManufacturerDräger
Ventilation ModesSIMV
Flow ControlElectronic
Tidal Volume Range20 to 1400 mL (model dependent)
Respiratory Rate4 to 60 breaths/min
PEEP0-20 cmH2O
Gas SupplyO2, Air, N2O
Vaporizer Mounts2
MonitoringO2, CO2, N2O
DimensionsVaries by model