Alarms
150 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
SpO2
Increment: 1
On Upper: 21 to 100%
Lower: 20 to 99%
100% (adult,
pediatric)
95% (neonate)
85% Off
NOTE:
Changing
this set-
ting auto-
matically
changes
the
Desat.
archive
setting to
the same
setting.
Desat.
Increment: 1
(Adult/
pediatric)
On
(neonate)
Lower: 19 to (SpO2 lower
limit - 1)% (within the
maximum measurement
range of 19 to 98%)
If the SpO2 low alarm limit
is lowered to or below the
SpO2 desat alarm limit,
then the SpO2 desat alarm
automatically adjusts to
1% less than the SpO2 low
alarm limit. The SpO2
desat alarm limit cannot
be raised to or above the
SpO2 low alarm limit
N/A 75% Off
NOTE:
Changing
this set-
ting auto-
matically
changes
the
SpO2
archive
setting to
the same
setting.
SpHb / SpHbv
Increment
0.2 g/dL
(0.1 mmol/L)
Upper: 1.2 to 25.0 g/dL
(0.7 to 15.5 mmol/L)
Lower: 1.0 to 24.8 g/dL
(0.6 to 15.4 mmol/L)
17.0 g/dL
(10.6 mmol/L)
7.0 g/dL
(4.3 mmol/L)
Off
PVI
Increment: 1
Upper: 1 to 100
Lower: 0 to 99
100 0 Off
SpCO
Increment: 1
Upper: 1 to 99
Lower: 0 to 98
10 0 Off
SpMet
Increment: 0.1
Upper: 0.1 to 99.9
Lower: 0.0 to 99.8
3.0 0 Off
Parameter Alarm
default
status
Alarm limit range Upper limit
defaults
Lower limit
defaults
Archive
default
setting