Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 151
Alarms
NIBP S adult
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 250 mmHg
1.5 to 33.3 kPa
Lower: 10 to 249 mmHg
1.3 to 33.2 kPa
160 mmHg
21.3 kPa
90 mmHg
12.0 kPa
Off
NIBP S pediatric
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 170 mmHg
1.5 to 22.7 kPa
Lower: 10 to 169 mmHg
1.3 to 22.6 kPa
120 mmHg
16 kPa
50 mmHg
6.7 kPa
Off
NIBP S neonate
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 130 mmHg
1.4 to 17.3 kPa
Lower: 10 to 129 mmHg
1.3 to 17.2 kPa
80 mmHg
10.7 kPa
50 mmHg
6.7 kPa
Off
NIBP D adult
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 250 mmHg
1.4 to 33.3 kPa
Lower: 10 to 249 mmHg
1.3 to 33.2 kPa
110 mmHg
14.7 kPa
50 mmHg
6.7 kPa
Off
NIBP D pediatric
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 170 mmHg
1.4 to 22.7 kPa
Lower: 10 to 169 mmHg
1.3 to 22.6 kPa
80 mmHg
10.7 kPa
35 mmHg
4.7 kPa
Off
NIBP D neonate
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 130 mmHg
1.4 to 17.3 kPa
Lower: 10 to 129 mmHg
1.3 to 17.2 kPa
60 mmHg
8kPa
25 mmHg
3.3 kPa
Off
NIBP M adult
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 250 mmHg
1.4 to 33.3 kPa
Lower: 10 to 249 mmHg
1.3 to 33.2 kPa
125 mmHg
16.7 kPa
60 mmHg
8.0 kPa
Off
Parameter Alarm
default
st
atus
Alarm limit
range Upper limit
defaults
Lower limit
defaults
Archive
default
setting