Alarms
152 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
NIBP M pediatric
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 170 mmHg
1.4 to 22.7 kPa
Lower: 10 to 169 mmHg
1.3 to 22.6 kPa
85 mmHg
11.3 kPa
40 mmHg
5.3 kPa
Off
NIBP M neonate
Increment:
1 mmHg or
0.1 kPa
On Upper: 11 to 130 mmHg
1.4 to 17.3 kPa
Lower: 10 to 129 mmHg
1.3 to 17.2 kPa
70 mmHg
9.3 kPa
40 mmHg
5.3 kPa
Off
ΔTx
Increment:
0.1 °C or
0.1
± 0.2 °F
Upper: 0.1 to 39.0 °C
0.2 to 70.2 °F
Lower: 0.0 to 38.9 °C
0.0 to 70.0 °F
1.0 °C
3.6 °F
0.0 °C
0.0 °F
Off
Txa/b
Increment:
0.1 °C
or 0.1 °F
Upper: 0.1 to 50.0 °C
32.2 to 122.0 °F
Lower: 0.0 to 49.9 °C
32.0 to 121.8 °F
39.0 °C
102.2 °F
34.0 °C
93.2 °F
Off
IBP S adult
Increment:
1 mmHg or
0.1 kPa
(GP1 S to
GP4 S,
LV S,
RV S)
On
(ART S,
PA S )
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
160 mmHg
(21.3 kPa)
for GP1 S to
GP4 S,
ART S, LV S
35 mmHg
(4.7 kPa) for
PA S, RV S
–90mmHg
(12.0 kPa)
for GP1 S
to
GP4 S,
ART S
–75mmHg
(10.0 kPa)
for LV S
–10mmHg
(1.3 kPa)
for PA S,
RV S
Off
Parameter Alarm
default
status
Alarm limit range Upper limit
defaults
Lower limit
defaults
Archive
default
setting