Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 153
Alarms
IBP S pediatric/
neonate
Increment:
1 mmHg or
0.1 kPa
(GP1 S to
GP4 S,
LV S,
RV S)
On
(ART S,
PA S)
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
–120mmHg
(16.0 kPa)
for GP1 S to
GP4 S,
ART S, LV S
–35mmHg
(4.7 kPa) for
PA S, RV S
50 mmHg
(6.7 kPa)
for GP1 S
to
GP4 S,
ART S
75 mmHg
(10.0 kPa)
for LV S
10 mmHg
(1.3 kPa)
for PA S,
RV S
Off
IBP D adult
Increment:
1 mmHg
or 0.1 kPa
(GP1 D to
GP4 D,
LV D,
RV D)
On
(ART D,
PA D)
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
–110mmHg
(14.7 kPa)
for GP1 D to
GP4 D,
ART D
–25mmHg
(3.3 kPa) for
LV D
–13mmHg
(1.7 kPa) for
PA D, RV D
50 mmHg
(6.7 kPa)
for GP1 D
to
GP4 D,
ART D
–2mmHg
(0.3 kPa)
for PA D,
LV D,
RV D
Off
IBP D pediatric
Increment:
1 mmHg
or 0.1 kPa
(GP1 D to
GP4 D,
LV D,
RV D)
On
(ART D,
PA D)
Upper: –24 to +300 mmHg
–3.2 to +40.0 kPa
Lower: –25 to +299 mmHg
–3.3 to +39.9 kPa
–80mmHg
(10.7 kPa)
for GP1 D to
GP4 D,
ART D
–25mmHg
(3.3 kPa) for
LV D
–13mmHg
(1.7 kPa) for
PA D, RV D
30 mmHg
(4.0 kPa)
for GP1 D
to
GP4 D,
ART D
–2mmHg
(0.3 kPa)
for PA D,
LV D,
RV D
Off
Par
ameter A
larm
default
status
Alarm limit range Upper limit
defaults
Lower limit
defaults
Archive
default
setting