Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 157
Alarms
inCO2
Increment of
1 mmHg,
0.1 kPa, or 0.1
On Upper: 2 to 10 mmHg
0.3 to 1.3 kPa
0.3 to 1.3%
Lower: not user-selectable
4 mmHg
(0.5 kPa, 0.5%)
Not applicable Off
etCO2
Increment of
1 mmHg,
0.1 kPa, or 0.1%
On Upper: 6 to 99 mmHg
0.8 to 13.3 kPa
0.8 to 13.3%
Lower: 5 to 99 mmHg
0.7 to 13.2 kPa
0.7 to 13.2%
50 mmHg
(6.7 kPa, 6.6%)
30 mmHg
(4.0 kPa,
3.9%)
Off
PVC/min
Increment of 1
On Upper: 1 to 50 10 Not applicable Off
inHal
Increment of
0.1 kPa or 0.1%
On Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%inHal
1.6 kPa, 1.6%
(adult)
1.9 kPa, 1.9%
(pediatric)
0.0 kPa, 0.0% Store
etHal
Increment of
0.1 kPa or 0.1%
Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%
8.5 kPa, 8.5% 0.0 kPa, 0.0% Off
inIso
Increment of
0.1 kPa or 0.1%
On Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%
2.4 kPa, 2.4%
(adult)
2.8 kPa, 2.8%
(pediatric)
0.0 kPa, 0.0% Store
Parameter Alarm
default
status
Alarm limit range Upper limit
defaults
Lower limit
defaults
Archive
default
setting