Alarms
158 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
etIso
Increment of
0.1 kPa or 0.1%
Upper: 0.1 to 8.5 kPa
0.1 to 8.5%
Lower: 0.0 to 8.4 kPa
0.0 to 8.4%
8.5 kPa, 8.5% 0.0 kPa, 0.0% Off
inEnf
Increment of
0.1 kPa or 0.1%
On Upper: 0.1 to 10.0 kPa
0.1 to 10.0%
Lower: 0.0 to 9.9 kPa
0.0 to 9.9%
3.6 kPa, 3.6%
(adult)
4.1 kPa, 4.1%
(pediatric)
0.0 kPa, 0.0% Store
etEnf
Increment of
0.1 kPa or 0.1%
Upper: 0.1 to 10.0 kPa
0.1 to 10.0%
Lower: 0.0 to 9.9 kPa
0.0 to 9.9%
10 kPa, 10% 0.0 kPa, 0.0% Off
inSev
Increment of
0.1 kPa or 0.1%
On Upper: 0.1 to 10.0 kPa
0.1 to 10.0%
Lower: 0.0 to 9.9 kPa
0.0 to 9.9%
4.4 kPa, 4.4%
(adult)
5.1 kPa, 5.1%
(pediatric)
0.0 kPa, 0.0% Store
etSev
Increment of
0.1 kPa or 0.1%
Upper: 0.1 to 10.0 kPa
0.1 to 10.0%
Lower: 0.0 to 9.9 kPa
0.0 to 9.9%
10 kPa, 10% 0.0 kPa, 0.0% Off
Parameter Alarm
default
status
Alarm limit range Upper limit
defaults
Lower limit
defaults
Archive
default
setting