Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 267
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
Show parameters
(middle button)
SpHb
1)
SpOC (default)
PVI
–SpCO
SpMet
Selects the parameter for the
parameter 2 location in the
Pulse CO-Ox parameter field.
With an Hb sensor, the default
parameter is SpOC. With a CO-
sensor, the default parameter
for the parameter 2 location in
the parameter field changes
automatically to SpMet.
Changes to the parameter
selection are retained if the
same sensor is disconnected
and then reconnected. The
parameter selection changes to
the default selection, if another
Masimo rainbow SET sensor
type is connected.
Show parameters
(right button)
SpHb
1)
–SpOC
PVI (default)
SpMet
Selects the parameter for the
parameter 3 location in the
Pulse CO-Ox parameter field.
PVI is the default parameter for
the parameter 3 location in the
parameter field for both CO and
Hb sensors.
Changes to the parameter
selection are retained if the
same sensor is disconnected
and then reconnected. The
parameter selection changes to
the default selection, if another
Masimo rainbow SET sensor
type is connected.
1)
Note: if the venous blood source was selected for SpHb Cal, the parameter label changes from SpHb
(arterial blood source) to SpHbv.
Selection Available settings Description