SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
268 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
SpHb averaging
time
For SpHb
1)
the selections are:
Long – approximately 6 minutes
Medium (default) – approximately 3
minutes
Short approximately 1 minute
Determines how responsive the
monitor is to rapid physiological
changes while tracking blood
hemoglobin values.
A longer averaging time
provides a more accurate
result. However, in clinical
situations where rapid
physiological changes have to
be monitored, use a shorter
averaging time.
Pulse CO-Ox mini
trend
–SpHb
1)
(default)
–SpCO
–SpOC
–SpMet
PVI
(SpCO is the default when a CO sensor is
used)
Selects the parameter to be
included in the mini-trend
display.
Color Red, green, blue, yellow, light blue, purple,
orange, white (default).
Determines the color of the
parameter labels and values.
1)
if the venous blood source was selected for SpHb Cal, the parameter label changes from SpHb
(arterial blood source) to SpHbv.
Selection Available settings Description