Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 557
Troubleshooting
Low %0 sensor failure The Agent sensor
measurement has failed due
to:
Sample line occlusion.
Electrical disturbance.
Internal failure.
Check sample line.
Remove radiating
devices (e.g.,
telephone).
Use alternative agent
measurement
system.
Call DrägerService
®
Low %0 value temporarily
unavail.
Agent parameter has
unknown accuracy or
automatic identification is
taking more time than usual,
possibly due to:
Zeroing failure
Polluted ambient air
during zeroing.
Electromagnetic distur-
bances.
Overheating.
Ensure clean ambi-
ent air.
Remove radiating de-
vices (e.g., tele-
phone).
Check ambient tem-
perature.
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Power cycle Scio
Change vaporizer
settings.
Call DrägerService
®
.
Low Check water
trap/sample line
Sample line is blocked or
not connected.
Water trap is full or not
installed.
Check sample line.
Check water trap.
Alarm Priority Alarm Cause Remedy