Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 557
Troubleshooting
Low %0 sensor failure The Agent sensor
measurement has failed due
to:
– Sample line occlusion.
– Electrical disturbance.
– Internal failure.
– Check sample line.
– Remove radiating
devices (e.g.,
telephone).
– Use alternative agent
measurement
system.
– Call DrägerService
®
Low %0 value temporarily
unavail.
Agent parameter has
unknown accuracy or
automatic identification is
taking more time than usual,
possibly due to:
– Zeroing failure
– Polluted ambient air
during zeroing.
– Electromagnetic distur-
bances.
– Overheating.
– Ensure clean ambi-
ent air.
– Remove radiating de-
vices (e.g., tele-
phone).
– Check ambient tem-
perature.
– Check water trap and
sample line.
– Change the water
trap or sample line if
necessary.
– Power cycle Scio
– Change vaporizer
settings.
– Call DrägerService
®
.
Low Check water
trap/sample line
– Sample line is blocked or
not connected.
– Water trap is full or not
installed.
– Check sample line.
– Check water trap.
Alarm Priority Alarm Cause Remedy