Troubleshooting
558 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low CO2 reduced
accuracy
Accuracy of the CO2 sensor
cannot currently be
guaranteed.
Ensure clean ambi-
ent air
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Wait for automatic ze-
roing
Power cycle the gas
analyzer.
Call DrägerService
®
.
Low %0 sensor failure The CO
2 sensor in patient gas
measurement module has
failed due to:
Sample line occlusion.
Electrical disturbance.
Internal failure.
Check sample line.
Remove radiating
devices (e.g.,
telephone).
Use alternative CO
2
measurement
system.
Call DrägerService
®
.
Medium %0 out of range high CO
2 concentration has
exceeded the Scio upper limit
of the measurement range.
Check vaporizer, fresh-
gas settings and ventila-
tion.
Low etAgent < #
Note
This alarm only occurs
for the primary agent.
Expiratory anesthetic gas
concentration has fallen
below the lower alarm limit
for more than 15 seconds.
Soda lime is dried out.
Check vaporizer and
fresh-gas settings.
Check breathing
system for large
leaks.
Exchange soda lime.
Medium etAgent > #
Note
This alarm only occurs
for the primary agent.
Expiratory anesthetic gas
concentration has exceeded
the upper alarm limit for more
than 15 seconds.
Check vaporizer and
fresh-gas settings
Medium etCO
2 < # Expiratory CO2 has fallen be-
low the limit for more than 15
seconds.
Check ventilation.
Alarm Priority Alarm Cause Remedy