Troubleshooting
560 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low Gas sensor reduced
accuracy
Accuracy of the gas
measurements cannot
currently be guaranteed.
Ensure clean
ambient air.
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Wait for automatic
zeroing
Power cycle the gas
analyzer.
Call DrägerService
®
.
Low inAgent < #
NOTE
This alarm occurs only
for the primary agent.
Inspiratory anesthetic gas
concentration has fallen
below the lower alarm limit
for more than 15 seconds
Soda lime is dried out
Check vaporizer and
fresh-gas settings.
Check breathing
system for large
leaks.
Exchange soda lime.
Medium inAgent > #
NOTE
This alarm occurs only
for the primary agent.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit:
At least 15 seconds (with
respiratory phases).
At least 30 seconds
(without respiratory
phases).
Check vaporizer and
fresh-gas settings.
Medium inN2O > 82% Inspired N2O is greater than
82%
At least 15 seconds (with
respiratory phases).
At least 30 seconds
(without respiratory
phases).
Check fresh-gas
composition.
Alarm Priority Alarm Cause Remedy