Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 561
Troubleshooting
Medium inCO2
> # Inspired CO2 has exceeded
the limit for more than 15 sec-
onds possibly due to one of
the following:
Soda lime is depleted.
Leakage in breathing sys-
tem.
Gas measurement is inac-
curate due to high respira-
tory rate.
Large dead space.
Check soda lime.
Increase fresh-gas
flow.
Check fresh-gas set-
tings.
Replace the breath-
ing system.
Adjust alarm limits if
necessary.
Check ventilation set-
tings.
Medium
Inspiratory xMAC
high
The inspiratory anesthetic gas
concentration has exceeded 3
xMAC for more than 3
minutes.
Check vaporizer and
fresh-gas settings.
High
Inspiratory xMAC
high
The inspiratory anesthetic
gas concentration has
exceeded 5 xMAC
or, while the patient is breath-
ing:
The inspiratory anesthetic
gas concentration has
exceeded 3 xMAC for
more than 30 seconds,
The expiratory anesthetic
gas concentration has
exceeded 2.5 xMAC for
more than 30 seconds.
Check vaporizer and
fresh-gas settings.
Medium %0 out of range high N2O concentration has
exceeded the Scio upper limit
of the measurement range.
Check vaporizer, fresh-
gas settings and ventila-
tion.