Troubleshooting
562 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low %0 reduced accuracy Accuracy of the N2O sensor
cannot currently be
guaranteed.
Ensure clean ambi-
ent air
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Wait for automatic ze-
roing.
Power cycle the gas
analyzer.
Call DrägerService
®
.
Low %0 sensor failure The N
2O sensor in patient gas
measurement module has
failed due to:
Sample line occlusion.
Electrical disturbance.
Internal failure.
Check sample line.
Remove radiating
devices (e.g.,
telephone).
Use alternative N
2O
measurement
system.
Call DrägerService.
Low %0 value temporarily
unavail.
N
2O parameter has unknown
accuracy possibly due to:
Zeroing failure.
Polluted ambient air
during zeroing.
Electromagnetic distur-
bances.
Overheating.
Ensure clean ambi-
ent air.
Remove radiating de-
vices (e.g., tele-
phone).
Check ambient tem-
perature.
Check water trap and
sample line.
Change the water
trap or sample line if
necessary.
Power cycle the gas
analyzer.
Call DrägerService
®
.
Alarm Priority Alarm Cause Remedy