Instruction Manual Dräger MSI EM200-E
Edition 2012-05-14 26 of 29 5695022
12. Error Messages and Operation References
After being switched on, as well as during the measurement process, the MSI EM200-E
checks the function of all measuring channels (see 3.3). Error messages are shown be-
fore the selection of functions or during normal function.
12.1 Error Messages
12.1.1 Error Messages Regarding Sensors
Error Message Error Cause
Remedy
CO-Sensor
Probe has been in flue during zero calibration
Sensor defect
Calibrate again with fresh air
Service
NO-Sensor
NO-Sensor has been without current
Probe has been in flue during zero calibration
Sensor defect
check function
see CO-sensor
NO2-Sensor
Probe has been in flue during zero calibration
Sensor defect
Calibrate again with fresh air
Service
Pressure Sensor Sensor defect
Service
12.1.2 General Error Messages
Error Message Error Reason; Remedy
System temperature operating temperature range exceeded;
vary temperature, service
Battery temperature
operating temperature range exceeded;
vary temperature, service
Battery voltage voltage not between 4.4 V and 6.5 V; replace battery (service)
Battery current battery current for charging or operating too high; service
Battery error battery manager data error; charge battery
Set clock clock lost data; charge battery, set clock
Options
data record options error; factory service
Settings
wrong settings; check and change settings
Calibration data
error in calibration data record; service
Data memory
error in memory data record; service
Next service
next service data error; service
System configuration
system configuration error, service