Estimate for expected time of operation of
Oxylog 2000
Example:
CMV mode, ventilator rate 10 bpm, VT = 1 L
minute ventilation = 10 bpm x 1 L = 10 L/min
Time of operation =
Time of operation = approx. 56 minutes
NOTE: Gas consumption is reduced by approximately
50 % and the time for operation increases to approxima-
tely 110 minutes when Oxylog 2000 is switched to
"Air Mix".
Supply with oxygen from wall pipeline
Attach high pressure hose with DISS connector to
Oxylog 2000 and connect to wall terminal with
appropriate quick connect probe.
Supply from O2-air blender
Note supply pressures for O2 and medical gas input:
40 to 88 psi (4.0 to 6.0 bar)
Carrying system 2000
The carrying system 2000 is recommended for quick
mobile use of the Oxylog 2000:
Accommodates the Oxylog 2000 ventilator, O2 cylinder
with pressure regulator, accessory bag and circuit holder
– for immediate use of the ventilator,
see "Ordering Information".
__________
* Gas consumption of ventilator: approximately 1 L/min
Preparation
Connecting Gas Supply
Operating Instructions Oxylog 2000, SW 3.n
22
Medical gas supply [L]
(MV +1*) [L/min]
622
11
03329033

Table of Contents

Other manuals for Dräger Oxylog 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 and is the answer not in the manual?

Dräger Oxylog 2000 Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals