192 Upute za uporabu | Dräger Polytron
®
6100 EC WL, Dräger Polytron
®
Repeater ISA
hr | Otklanjanje smetnji
11.1 ISA100 Wireless
standardni objekti
11.2 Status mjerne vrijednosti (PV_STATUS)
Znaenje „Quality” (Bit 6 i 7):
0: Loš (Bad) – nema prijenosa mjerne vrijednosti.
1: Nesiguran (Uncertain) – kvaliteta mjerenja manja je od uobiajene.
2: Dobar (Good) – kvaliteta mjerenja je dobra, ali bi mogao postojati uvjet alarma.
Prikaz Pristup Oktet br. Sadržaj Ulazni registar Mod-
bus-a
Tip podataka
PV READ 1 PV_Status Register N UINT (16 bit)
PV READ 2 PV_Value (1/4)
Register N+1
32-Bit-Float (4 okteta),
Float32 prema IEEE 754
PV READ 3 PV_Value (2/4)
PV READ 4 PV_Value (3/4)
Register N+2
PV READ 5 PV_Value (4/4)
DIAG_STATUS READ 1 Diag_Status (1/4)
Register M UINT (16 bit)
DIAG_STATUS READ 2 Diag_Status (2/4)
DIAG_STATUS READ 3 Diag_Status (3/4)
Register M+1 UINT (16 bit)
DIAG_STATUS READ 4 Diag_Status (4/4)
Quality Substatus
Bit 6 i 7 Bit 2 do 4 PV Uzrok Pomoć
0 0 NaN Aktivna je poetna sekvenca. Priekajte poetnu sekvencu.
Poetna sekvenca je neuspješna, EC mjerna
glava nije prepoznata.
A Ponovno pokrenite ureaj za mjerenje plinova.
BUreaj za mjerenje plinova povjerite na pro-
vjeru tvrtki DrägerService.
0 1 NaN Aktivirana je funkcija zakljuavanja senzora.
Umetnut je senzor s razliitim brojem artikla.
Upotrijebite senzor s istim brojem artikla ili deakti-
virajte funkciju zakljuavanja senzora.