Otklanjanje smetnje | sr
Uputstvo za korišćenje | Dräger Polytron
®
6100 EC WL, Dräger Polytron
®
Repeater ISA 267
11.1 ISA100 Wireless
standardni objekti
11.2 Status izmene vrednosti (PV_STATUS)
Znaenje "Quality" (Bit 6 i 7):
0: Loše (Bad) – bez prenosa merne vrednosti.
1: Nesigurno (Uncertain) – kvalitet merenja vrednosti je niži nego uobiajeno.
2: Dobar (Good) – kvalitet merenja vrednosti je dobar, ali može postojati uslov za alarm.
Prikaz Pristup Oktet br. Sadržaj Modbus registar unosa Tip podataka
PV READ 1 PV_Status Register N UINT (16 bita)
PV READ 2 PV_Value (1/4)
Register N+1
32-Bit-Float (4 okteta),
Float32 prema IEEE 754
PV READ 3 PV_Value (2/4)
PV READ 4 PV_Value (3/4)
Register N+2
PV READ 5 PV_Value (4/4)
DIAG_STATUS READ 1 Diag_Status (1/4)
Register M UINT (16 bita)
DIAG_STATUS READ 2 Diag_Status (2/4)
DIAG_STATUS READ 3 Diag_Status (3/4)
Register M+1 UINT (16 bita)
DIAG_STATUS READ 4 Diag_Status (4/4)
Quality Substatus
Bit 6 i 7 Bit 2 do 4 PV Uzrok Rešenje
0 0 NaN Aktivan redosled pokretanja. Saekajte redosled pokretanja.
Redosled pokretanja nije uspeo, nije prepoznata
EC merna glava.
A Pokrenite ponovo ureaj za merenje gasa.
B Neka vam ureaj za merenje gasa proveri
DrägerService.
0 1 NaN Funkcija blokade senzora je aktivirana. Korišen
je senzor s brojem dela koji odstupa.
Koristite senzor s istim brojem dela ili deaktivirajte
funkciju blokade senzora.