41
DRAMIŃSKI BLUE – user manual
10.6. Optimization of B+M and M modes
10.6.1. Graph drawing speed selection in M mode
In order to activate B+M mode press [B+M Mode] button. The submenu will appear over the button, where you
can select graph drawing speed in M mode. Available values: 1s, 2s, 3s, 4s.
10.6.2. Setting the sampling line in B+M / M modes
Setting the sampling line is based on B mode image. In order to change its position drag and drop it with your nger.
10.6.3. Switching between M and B+M modes
After pressing [B+M mode] button new submenu will appear, where on the top there is [M/B+M] button. It is
used for switching between B+M and M modes.
10.7. Imaging with the function of color Doppler
The Color Doppler functions allow viewing a blood ow in a real time together with B Mode image in the
background. The blood ow is coded by the color depending on its direction. The red color corresponds to the
stream owing to the head of the probe. Blue color means the reverse direction.
Attention! By double click on color bar reverse colors option is activated (stream owing to the head of the
probe will be coded with blue color, red color means the reverse direction).
10.7.1. Starting Color Doppler imaging
Remember that preparing patient to examination has great inuence on results. Use of an appropriate amount
of gel is essential for obtaining a good result.
Before starting Doppler function you should nd the clearest image of the organ in B-Mode, showing the ves-
sels. Too high Gamma and Gain setting the image in B-Mode will have a negative inuence on the quality of
the Doppler image.
In order to start this mode you should click [Doppler] button in the mode panel.
The screen will show the active panel managing the Color Doppler function and Doppler frame.
10.7.2. Switching between the Doppler modes: Color, Power, Pulse Wave; Doppler mode OFF
To switch between Doppler modes use [Color Doppler], [Power Doppler] and [Pulse Doppler] buttons. Theyare
visible under the [Gamma] button after pressing [Doppler] button.
In order to nish work in Doppler mode press [Doppler] button again. The device will automatically return to
the B Mode.
Color Doppler = Doppler with color coding of the blood vessel ow
Power Doppler = Doppler with coding the power of the ow
Pulse Doppler = Doppler used to measure the ow speed