DRAMIŃSKI BLUE – user manual
11.6. Audio
You can adjust the volume of sound in Audio tab.
To check the sound volume, use the [Test sound] option.
11.6.1. Silent mode
Activating silent mode will mute all message sounds.
To activate silent mode, go to [Settings] and select the [Audio] tab.
Select the [Silent Mode] check box.
Uncheck the box to deactivate it.
Warning! In the silent mode, the system does not emit any acoustic signals, including information on the state
of battery charge or its deep discharge.
11.7. Measurement settings
The [Measures] tab is used to modify the list of displayed measurements in the age tables.
To turn o displaying of the measurements clear the check box situated next to its name.
11.8. DICOM settings
The DICOM settings window can be accessed in two ways:
1. Press [Settings] -> [DICOM], or
2. In the patient database, expand the list [Send DICOM] -> [DICOM Settings]
11.8.1. DICOM servers conguration
Warning! Make sure the ultrasound scanner is properly connected to the facility's internal network or the In-
ternet using a LAN cable or an external WiFi antenna (accessory).
To congure data export to the DICOM server, go to [Settings] and select the [DICOM] tab.
In the DICOM Record eld, press [Add].
Fill in the elds according to the information from the server provider.
Specify whether the server should be treated as default using the [Default] check box.
Select the encoding method appropriate for the server from the [Encoding] list.
In the [Transfer syntax] list, specify the compression method for the server.
From the [Transfer syntax conversion] list, select the allowable server conversion option.
To check if the connection with the server is correct press [Echo]. The Echo OK message means that the ultra-
sound machine has connected to the server.
To nish the conguration press [Save]. The server will appear in the list in the DICOM Write eld.