DRAMIŃSKI BLUE – user manual
2. User's safety
Warning!
The user's and the patient's safety depends on observing the below
mentioned instructions!
1. DRAMIŃSKI BLUE should be used for diagnostic purposes and by the trained sta only – erinarians pro-
cient in ultrasound diagnostics.
2. It is necessary to disinfect the probe before each use. Other elements of the ultrasound scanner should be
disinfected in reasonable situations in which they could have a contact with infectious substances.
3. It is forbidden to use the ultrasound scanner with high frequency (HF) devices at the same time.
4. The ultrasound scanner cannot be used for transesophageal echocardiography (TEE) examinations.
5. In the places where explosive and anaesthetic gases are used, the use of the ultrasound scanner is prohib-
ited for safety reasons.
6. The device is suitable for indoor use.
7. It is recommended the user of the ultrasound scanner perform regular technical inspections every two
years at the manufacturer’s facility or approved service point. It will guarantee the highest level of the
patient's safety.
8. It is forbidden to carry out ultrasound diagnostics when carrying out debrillation.
9. It is forbidden to disassembly the device, repair and adjust it by the user except for the procedures stated
in this manual.
10. It is recommended the users of ultrasound scanner check before the examination the condition of the
probe, cable and the connector. Do not use the probe if any of these elements is damaged.
11. If mechanical damages of the probe, cable or connector are noticed it is necessary to send the device to
the service point authorised by DRAMINSKI.
12. Even though the ultrasound scanner has a strong construction in order to avoid mechanical damages it is
recommended to follow the instructions presented in this manual.
13. Exposing this device to strong solar radiation should be avoided. Regarding working, storage and trans-
port conditions, it is advised to follow the instructions on the labels of the device and its parts.
14. It is forbidden to modify the device by the user.
15. DRAMIŃSKI BLUE is an electric device which can be a source of electromagnetic radiation. Other electric
devices can interfere with its work, which is why it is recommended to limit the number of other electric
devices working nearby.
16. When its service life expires in order to avoid risk to the environment, the device and the accessories
should be disposed of by specially trained units according to the applicable regulations or sent back to
the manufacturer.
17. During the ultrasound examination the device cannot be connected to the ETHERNET (via LAN connector).
18. It is recommended to connect medical grade external monitors and printers.
19. In order to avoid electric shock the device should be connected to a grounded outlet.