The AnyDesk window with the customer number will appear on the screen. Provide consultant with this num-
ber.
Wait for the consultant to take over the ultrasound scanner control. It should take several seconds.
Warning! Remember not to turn o the camera, router and Internet connection while the consultant is carry-
ing out service work using remote access. This may result in system malfunction!
11.3.3.2. Disabling remote access
To disable remote access, go to [Settings] and select the [Information] tab.
Press [Remote Access STOP].
You can also disable remote access by interrupting the ultrasound machine's connection to the Internet.
11.3.4. Security lock in the operation system
To unlock the possibility of introducing changes to the system, go to Settings-> Information and uncheck the
"Secure operating system" box. Conrm the system restart in the dialog box
11.4. Date and time settings
Enter [Settings]. /Time/ tab is used to update system date and time.
Enter actual date and time. Next click [Save date and time] button.
11.5. Display
In section /Display/ brightness and touch panel calibration options are available.
11.5.1. Touch screen calibration
BLUE Ultrasound scanner is calibrated.
In case of system breakdown, it may be necessary to calibrate the touch panel again. It is when the sensitivity
and accuracy of the panel is lower.
To calibrate the touch panel got to [Settings] -> [Display]. Next tap [Touch panel calibration].
1. eGalaxTouch settings window will appear.
2. Go to [Tools] tab.
3. Perform the “Draw Test”. Try to draw with your nger the lines overlaying the grid and assess the accuracy
of the panel.
4. If the touch panel reacts incorrectly, choose one of the options – 4 – and 9-points calibration.
5. Follow the tips displayed on the screen.
If you can not use the touch screen to enter the calibration mode, connect and use mouse to do that.