34
Drive recommends the following maintenance and cleaning procedures be conducted between users.
ELECTRONICS
Check all controls to make sure all functions work properly.
Foot control
Head control
Check all cables for damaged or frayed wires.
Power cord
Pendant cord
Check to make sure all plugs are fully inserted or attached.
Check 9V battery on pendant and motor and replace annually.
BED FRAME & SLEEPING SURFACE
Visually check all welds.
Head section
Foot section
Main Frame
Check joints between sleeping surface sections for loose fasteners.
CLEANING
If possible, remove patient/resident before cleaning bed.
Prior to cleaning unplug power supply cord
Ensure all electrical parts (motors, control boxes, wires and pendant) are not broken. Ensure that NO liquids enter
electrical components.
The metal parts of the bed are covered with a powder coating. Clean all coated parts with mild detergent and warm
water.
Raise head and foot sections of the bed to remove dust from frame.
Remove mattress and clean mattress deck.
Remove all gross/solid contaminants, then wash and sanitize all components. DO NOT submerge the bed frame or
electrical components.
Solvents, alcohol, bleach, caustic agents, high acid or alkaline solutions or petroleum based products should not be
used on the bed.
Use standard water pressure. DO NOT power wash or steam clean any parts.
Rinse completely with water (Maximum temperature 110
o
F or 43
o
C). Do not rinse with cold water.
Ensure all parts are dry before using or storing.
NOTE: This bed is NOT compatible with wash down tunnels.
LUBRICATION & MECHANICAL
Lubricate all caster roller and swivel bearings with light machine oil.
Check all bolts and tighten as needed.
WARRANTY
Lifetime warranty on welds and frame.
5 year warranty on motor and hand control.
1 year warranty on all other parts and components.
During the warranty period, defective items will be repaired or replaced at Drive’s option at no charge.
NOTE: The Competitor II bed can be used with a free standing trapeze or a Drive Medical frame mounted trapeze
(Item# 13009KT-BV).