D-FL 200 / Rev. 6 39
______________________________________________________________________________________________________
Industrie Elektronik GmbH & Co KG
Measuring head
: No. length inside the flange complete length
1 260 mm / 10,2 in 470 mm / 18,5 in
2 360 mm / 14,2 in 570 mm / 22,4 in Standard
3 560 mm / 22,0 in 770 mm / 30,3 in
4 760 mm / 29,9 in 970 mm / 38,2 in
5 960 mm / 37,8 in 1170 mm / 46,1 in
Ö
To combine the measuring head and the flange pay attention that
measuring head no. and flange no. are equal !!
Flange:
Pipe 110x5 mm (4,3x0,2in)
Flange ∅ 190 mm (7,5in)
Material: 1.4571 (standard)
Flange length:
Flange-No. length of the flanges
1 235 mm / 9,2 in
2 335 mm / 13,2 in Standard
3 535 mm / 21,1 in
4 735 mm / 28,9 in
5 935 mm / 36,8 in
Ö
To combine the measuring head and the flange pay attention that
measuring head no. and flange no. are equal !!
Purge air unit “G” for gas above dew point
Mains: 230V,±10%, 50Hz Others on request
Power consumption: 0,25kW at 230V
Air volume at 0 mmWS: app. 1,4 m³/min (49,4ft³/min)
Weight: app. 20kg
Purge air unit “E“ for gas below dew point
Mains: 200...240V, ∆-Wiring
345...415V, Y-Wiring
50Hz
… 60 Hz
Other voltage and
frequencies on request
Nominal current: 2,1A ∆-Wiring, 1,2A Y-Wiring
Power consumption: 0,43 kW at 230V, 50Hz
Air volume at 0 mmWS: app. 2,2 m³/min (77,7ft³/min)
Weight: app. 30 kg
One purge air unit includes 10m (33ft) purge air hose.