Attach Scaler handpiece
Attach the Scaler handpiece to the hose con-
nection.
Inserting/changing instruments
CAUTION
If an attempt is made to perform a
treatment with a damaged or worn
instrument, this can cause injury and
may result in an unsuccessful treat-
ment.
Worn or bent instruments should be
replaced immediately and not reused!
The tool kit cover of the instrument is
used as a torque wrench for tightening
the union nut of the instrument chuck.
Select a suitable instrument for the corre-
sponding treatment.
To prevent the instrument from being
overtightened, the torque wrench slips
when the required torque is reached. No
ratcheting sound can be heard.
Always use the torque wrench to screw the
instrument on and off.
When screwing the instrument on, slowly turn
the torque wrench approx. a quarter turn
beyond the resistance point.