15 Maintenance
15.1 Changing the valve in the
fluid container
The valve on the underside of the fluid container
must be cleaned regularly and checked for
blockages and leaks.
Press the valve lightly against the fluid con-
tainer and unscrew it anti-clockwise.
Clean the valve.
If cleaning is not possible, e.g. if the filter in the
valve is blocked, then the valve needs to be
replaced.
Check the O-ring.
In the event of leaks, loose seating or visible
damage the O-ring must be replaced.
Insert the valve in the holder and tighten it
clockwise as far as it will go.
15.2 Checking instrument wear
Instrument wear can be checked using the Vec-
tor toolcard:
Place the handpiece against the toolcard. If
the instrument tip extends beyond the red
marking then the instrument can still be used.
If the instrument tip just reaches the red mark-
ing then the instrument shows signs of wear
but can still be used.
If the instrument tip does not reach the red
marking then the instrument must be replaced.