Error Possible cause Remedy
Orange "FUNCTION" LED
lights up or flashes
The treatment was interrupted
as water may be present in the
following areas:
between the handpiece and the
handpiece hose.
Clean the affected areas and
dry using the air and water
syringe.
Instrument contact pressure too
high during treatment.
Reduce the contact pressure
and press the flexible foot
switch, the LED will go out.
If necessary, the instrument
might require spraying and
drying using the air and water
syringe.
Instrument defective.
Replace the instrument.
Handpiece defective
Replace the handpiece. Send
in the defective handpiece for
repairs.
Orange "PEDAL" LED lights
up
The power of the battery in the
flexible foot switch is low.
Check battery performance
and insert a new battery if
necessary.
Orange "PEDAL" LED flashes,
No flexible foot switch connec-
ted (cable operation) or paired
(wireless operation).
Connect a flexible foot switch
(cable operation) or perform
pairing (wireless operation).
Blue "CLEAN" LED continues
to light up after a cleaning
program has been carried out.
The CLEAN function of the
cleaning process was not fully
completed, or it was stopped
early.
Carry out the cleaning proc-
ess CLEAN in full.
The blue "POWER" and "LIQ-
UID" LEDs fail to light up after
the unit is switched on.
Handpiece is not recognised by
the basic unit.
Attach a different handpiece in
position. If the "POWER" and
"LIQUID" continue to fail to
light up, call a Service Techni-
cian.
Handpiece is defective.
Attach a different handpiece in
position. If the "POWER" and
"LIQUID" continue to fail to
light up, call a Service Techni-
cian.
The illumination in the Scaler
handpiece becomes increas-
ingly dimmer.
The light conductor has become
opaque or has become milky.
Replace the light conductor.
Illumination LEDs are defective.
Send in the defective Scaler
handpiece for repairs.
Troubleshooting
9000-615-32/02 2012V003 49
EN