13.11 Issue clearance for the parts
for sterilisation
The reprocessing of the medical products ends
with the documented clearance for storage and
renewed use.
Document the clearance of the medical prod-
uct after reprocessing.
13.12 Storing parts for sterilisation
Comply with the stated storage conditions:
Store the parts protected against contami-
nation
Dust-protected, e.g. in a locked cabinet
Protected against moisture
Protected against excessive temperature
fluctuations
Protected against damage
Packaging for a sterile medical device can suf-
fer damage as a result of a particular incident
and the passage of time.
Potential external contamination of the sterile
barrier system should be taken into account in
terms of aseptic preparation when establishing
the storage conditions.
14 Treatment breaks for more
than 24 hours
If no treatment is carried out for a period of
24 hours or more, reprocessing of the hose sys-
tem must be performed.
14.1 Cleaning and disinfecting the
hose system
The hose system is disinfected using a ready-to-
use, aldehyde-free solution of Vector/RinsEndo
Disinfection.
Pull off the hose connection from the hand-
piece.
Lay the handpiece hose in the sink.
Rinsing with water:
Fill the fluid container to approx. 1/3 with
water.
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
RINSE
LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
Rinsing with water will flush out any remain-
ing blockage.