Check the Scaler handpiece light conductor for
its light transmission, replace if necessary.
13.5 Manual cleaning, intermedi-
ate rinsing, disinfection, final
rinsing, drying in the cleaning
and disinfection bath
A disinfectant or combined cleaning and disinfec-
tant agent is required for manual disinfection. It
must have the following properties:
certified, possibly virucidal efficacy (DVV/RKI,
VAH or European Standards)
For further information, see: "General informa-
tion".
Cleaning
Place the removable parts of the handpiece
(Scaler cover, light conductor), instrument hold-
ers of the tool kits (without instruments) and
the disassembled handpiece in the cleaning
bath for the required reaction time, so that all
parts are covered.
Clean all accessible external surfaces and inter-
nal surfaces below the surface for 5 minutes
with a hygienic cleaning brush until all visible
soiling has been removed.
Place instruments in small parts baskets into
the cleaning bath.
Rinse handpiece through at least 3 times using
a 20-ml disposable syringe.
20 ml
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
20 ml
Remove all rinse adapters.
Comply with the cleaning agents' action times
specified by the manufacturer.
Usage
9000-615-32/02 2012V003 35
EN