Intermediate rinsing
After the action time prescribed by the manufac-
turer:
Rinse all components under running water for
at least 1 minute (temperature < 20 °C).
Rinse handpiece through at least 3 times using
a 20-ml disposable syringe.
20 ml
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette.
20 ml
Disinfection
Place the removable parts of the handpiece
(Scaler cover, light conductor), instrument hold-
ers of the tool kits (without instruments) and
the disassembled handpiece in the cleaning
bath for the required reaction time, so that all
parts are covered.
Clean all accessible external surfaces and inter-
nal surfaces below the surface for 5 minutes
with a hygienic cleaning brush until all visible
soiling has been removed.
Place instruments in small parts baskets into
the cleaning bath.
Usage