The reprocessing method was validated as fol-
lows:
Pre-cleaning
FD 350 Disinfection wipes (Dürr Dental)
Cleaning brush
Manual cleaning
ID 215 Enzymatic instrument cleaner (Dürr
Dental)
Cleaning brush
Manual disinfection
ID 213 Instrument disinfection (Dürr Dental)
Automatic cleaning and disinfection
was performed in accordance with EN ISO
15883 with tested efficacy.
Cleaning agent: Neodisher MediClean Forte
Washer-disinfector: PG 8535 (Miele)
Programmes: "Cleaning without neutralisa-
tion" and "THERMAL DES"
Rinsing adapter: Miele 68551101 D
Cleaning brush
Steam sterilisation
was performed in accordance with EN ISO
17665 using the fractionated vacuum proce-
dure.
Pre-vacuum: 3 x
Sterilisation temperature: 132°C
Sterilisation time: 2 minutes (half-cycle)
Drying time: min. 20 minutes
Cleaning brush
Cleaning brush with nylon hairs, double-sided
Number of brush heads: 2
Brush material: nylon
Brush head length: 25 and 35 mm
Brush length: 5 and 10 mm
Example: Interlock cleaning brush double-
sided green REF 09098
General information
NOTICE
Equipment damage due to unsuitable
products
Oils and care products containing oil will
damage the device.
The handpiece must not be maintained
with oil or with a care system that con-
tains oil.
Comply with all national directives, standards
and specifications for the cleaning, disinfection
and sterilisation of medical products as well as
the specific specifications for dental practices
and clinics.
Comply with the specifications (see "13.6 Man-
ual cleaning, intermediate rinsing, disinfection,
final rinse, drying in ultrasonic bath" and "13.7
Automatic cleaning, intermediate rinsing, disin-
fection, final rinse, drying") when selecting the
cleaning and disinfectant agents to be used.
Comply with the concentration, temperature,
residence time and post-rinsing specifications
issued by the manufacturer of the cleaning and
disinfectant agent.
Only use cleaning agents that are non-fixing
and aldehyde-free and display material com-
patibility with the product.
Only use disinfectants that are aldehyde-free
and display material compatibility with the
product.
Do not use any rinse aid (danger of toxic resi-
due on the components).
Only use freshly-produced solutions.
Only use distilled or deionised water with a low
bacterial count (at least drinking water quality)
that is free from facultatively pathogenic micro-
organisms (e.g. legionella bacteria).
Use clean, dry, oil and particle-free com-
pressed air.
Do not exceed temperatures of 138 °C.
Subject all devices used (ultrasonic bath,
cleaning and disinfection device (CD), sealing
device, steam steriliser) to regular maintenance
and inspections.
Usage
9000-615-32/02 2012V003 33
EN