The reprocessing method was validated as fol-
lows:
– Pre-cleaning
– FD 350 Disinfection wipes (Dürr Dental)
– Cleaning brush
– Manual cleaning
– ID 215 Enzymatic instrument cleaner (Dürr
Dental)
– Cleaning brush
– Manual disinfection
– ID 213 Instrument disinfection (Dürr Dental)
–
Automatic cleaning and disinfection
was performed in accordance with EN ISO
15883 with tested efficacy.
– Cleaning agent: Neodisher MediClean Forte
– Washer-disinfector: PG 8535 (Miele)
– Programmes: "Cleaning without neutralisa-
tion" and "THERMAL DES"
– Rinsing adapter: Miele 68551101 D
– Cleaning brush
–
Steam sterilisation
was performed in accordance with EN ISO
17665 using the fractionated vacuum proce-
dure.
– Pre-vacuum: 3 x
– Sterilisation temperature: 132°C
– Sterilisation time: 2 minutes (half-cycle)
– Drying time: min. 20 minutes
–
Cleaning brush
Cleaning brush with nylon hairs, double-sided
– Number of brush heads: 2
– Brush material: nylon
– Brush head length: 25 and 35 mm
– Brush length: 5 and 10 mm
Example: Interlock cleaning brush double-
sided green REF 09098
General information
NOTICE
Equipment damage due to unsuitable
products
Oils and care products containing oil will
damage the device.
❯
The handpiece must not be maintained
with oil or with a care system that con-
tains oil.
❯
Comply with all national directives, standards
and specifications for the cleaning, disinfection
and sterilisation of medical products as well as
the specific specifications for dental practices
and clinics.
❯
Comply with the specifications (see "13.6 Man-
ual cleaning, intermediate rinsing, disinfection,
final rinse, drying in ultrasonic bath" and "13.7
Automatic cleaning, intermediate rinsing, disin-
fection, final rinse, drying") when selecting the
cleaning and disinfectant agents to be used.
❯
Comply with the concentration, temperature,
residence time and post-rinsing specifications
issued by the manufacturer of the cleaning and
disinfectant agent.
❯
Only use cleaning agents that are non-fixing
and aldehyde-free and display material com-
patibility with the product.
❯
Only use disinfectants that are aldehyde-free
and display material compatibility with the
product.
❯
Do not use any rinse aid (danger of toxic resi-
due on the components).
❯
Only use freshly-produced solutions.
❯
Only use distilled or deionised water with a low
bacterial count (at least drinking water quality)
that is free from facultatively pathogenic micro-
organisms (e.g. legionella bacteria).
❯
Use clean, dry, oil and particle-free com-
pressed air.
❯
Do not exceed temperatures of 138 °C.
❯
Subject all devices used (ultrasonic bath,
cleaning and disinfection device (CD), sealing
device, steam steriliser) to regular maintenance
and inspections.
Usage
9000-615-32/02 2012V003 33
EN