Scaler instrument P4, 120 mm, for the
removal of supragingival coatings on smooth
surfaces and for interdental areas.
POWER
2–3 LEDs: 40 % - 60 % operating perform-
ance
P4
11.4 After every treatment
End of treatment
Dismantle the instrument with the tool kit cover.
After every treatment, all parts used must be
cleaned, disinfected and, if necessary, steri-
lised, "13 Reprocessing".
12 Cleaning
12.1 Cleaning of the outside surfa-
ces
All outside surfaces must be cleaned and disin-
fected if they are contaminated or soiled.
Surface of the device
Handpiece hose
Fluid container
Tool kit for Scaler without instruments
As the surface disinfectant, we recommend using
a disinfectant that is compatible with the materi-
als and meets the general dental hygiene stand-
ards, e.g.:
Dürr Dental FD 322 rapid surface disinfectant
Dürr Dental FD 350 disinfection wipes
Dürr Dental FD 366 sensitive rapid surface dis-
infectant
NOTICE
Liquid can cause damage to the unit.
Do not spray the unit with cleaning and
disinfectant agents.
Make sure that liquid does not get
inside the unit.
For pre-cleaning, remove coarse organic soiling
with a paper towel.
Clean the surfaces with a moist, soft, lint-free
cloth.
Usage
28 9000-615-32/02 2012V003