Place instruments in small parts baskets into
the ultrasonic bath.
❯
Place the handpiece without covers in a con-
tainer with fluid. The drive mechanism of the
handpiece must not lie in the fluid (malfunc-
tion). Therefore observe instructions concern-
ing maximum fluid level for the Scaler hand-
piece.
❯
Place container with the handpieces in the
ultrasonic bath with a suitable carrier.
❯
Rinse handpieces through at least 3 times
using a 20-ml disposable pipette.
❯
Remove all rinse adapters.
❯
Screw Scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times using a
20-ml disposable syringe.
❯
Comply with the cleaning agents' action times
specified by the manufacturer.
Intermediate rinsing
After the action time prescribed by the manufac-
turer:
❯
Rinse all components under running water for
at least 1 minute (temperature < 20 °C).
❯
Rinse handpiece through at least 3 times using
a 20-ml disposable syringe.
❯
Screw scaler instruments onto the rinsing
adapter one after another and rinse through
each of the instruments at least 3 times with
water using a 20-ml disposable pipette.
Usage
38 9000-615-32/02 2012V003
EN