When the calibraon procedure begins ---- appears on the screen. Once Calibraon is complete
donE appears on screen. If there are any Faults with the calibraon an error message will appear
either Er_r or Er_C. Er_C means the low calibraon will fail if the oset is less than -10mV or greater
than +10mV. This signies potenal faulty sensors or the high calibraon will fail if the count value is
less than +20mV or greater than +50mV. This signies potenal faulty sensors. Er_r means the high
calibraon will fail if the mV value is within 10mV of the low calibraon value. This is a potenal
RCAL failure.
6. Messages & Error Indicaons
These messages indicate that the instrument may require aenon, or there is a problem with the
signal input connecon. The message legend is shown for 1 second, followed by its value.
Cauon: Do not connue with the process unl the issue is resolved.
Note: CHHJ, CLLJ or OPEN may be displayed if an incorrect input type is selected.