5.1.9 Decimal Point (DP)
In the setting we can select decimal point as per the scale capacity and
division.
0(No Decimal Point)
1(0.0)
2(0.00)
3(0.000)
4(0.0000)
5.1.10 Unit (UNT)
This is scale default unit setting. We can make the scale either in
kg, g or Lb.
kg
g
Lb
Press ESC key to go to the previous menu.
5.2 Factory Calibration (FCAL)
After setting the range, capacity, division, decimal point and unit, we
have press the Tare Key then the display shows FCAL (Factory
Calibration). Then follow the following process.
1) The display shows ZCAL (zero calibration).We have to make sure
there is no weight in the platform and then press the M+/MR Key (enter).
Now the display will show AD counts.
2) Once again press the M+/MR Key. The display will show HLOAD and
we have to place 50% of the full scale weight and then the display will
show the corresponding AD count after pressing the Enter key.