INTRAORAL RADIOGRAPHY EQUIPMENT
The Intraoral Dental Digital Imaging Sensor is compatible with intraoral x-ray
equipment with electronic timer that complies with current standards EN 60601-1, EN
60601-1-2, EN 60601-1-3 and IEC 60601-2 -65 and meets the following requirements:
The intraoral X-ray generator must operate at a voltage between 60 and 70 kV
and an anode current of 7mA.
The intraoral X-ray generator must be installed according to the manufacturer's
instructions and requirements.
Don't use generators with power of less than 60KV.