EasyManua.ls Logo

Eagle new IDA - Guidance and Declaration for Electromagnetic Immunity

Eagle new IDA
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
GUIDANCE AND DECLARATION FOR ELECTROMAGNETIC IMMUNITY
Phenomenon
Basic standard of
EMC
Immunity test level
Level of conformity
Electrostatic discharge
IEC 61000-4-2
±8 KV contact
± 2 KV, ± 4 KV, ± 8 KV,
± 15 KV air
±8 KV contact
± 2 KV, ± 4 KV, ± 8 KV,
± 15 KV air
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz 2.7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz 2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
See table
See table
Electrical Fast / Transients
bursts
IEC 61000-4-4
± 1 kV
100 kHz repetition
frequency
± 1 kV
100 kHz repetition
frequency
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V
0.15 MHz 80 MHz
6 V m in ISM bands
between 0.15MHz e
80MHz
80 % AM at 1 kHz
3 V
0.15 MHz 80 MHz
6 V m in ISM bands
between 0.15MHz e
80MHz
80 % AM at 1 kHz
Rated power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz

Table of Contents