ELECTROMAGNETIC COMPATIBILITY
The Intraoral Dental Digital Imaging Sensor is intended for use in the
electromagnetic environment specified below. Buyer or user should ensure that it is
used in such an environment.
The Intraoral Dental Digital Imaging Sensor is suitable for use in a
professional health care environment, not including areas where there are sensitive
equipment or sources of intense electromagnetic disturbances, such as the RF shielded
room of an imaging system magnetic resonance imaging, in operating rooms near
active AF surgical equipment, electrophysiology laboratories, armored rooms or areas
where short wave therapy equipment is used.
The following tables provide information on compliance of the equipment
according to the standard EN 60601-1-2:2015.
GUIDANCE AND DECLARATION FOR ELECTROMAGNETIC EMISSIONS
Emission Essays
Conformity
Electromagnetic environments -
guidelines
RF emissions CISPR 11
Group 1
The Intraoral Dental Digital Imaging
Sensor uses RF energy only for its
internal functions. Therefore, your RF
emissions are very low and probably
will not cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The Intraoral Dental Digital Imaging
Sensor is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuation/
Scintillation Emissions
IEC 61000-3-3
Not applicable