Warning "There's a difference
between the serial number of the
New IDA sensor and the
calibration files. If the calibration
files belongs to this sensor,
please ignore this message
Calibration file doesn't belong to the sensor
- Contact the technical service
Calibration file belongs to the sensor, but it was generated
with a wrong serial number.
- Ignore warning
Error "To run this application, it is
necessary to install one of the
following versions of .NET
Framework v4.0"
Fail to install.NET 4.5.1.
- Install Microsoft.NET 4.5.1 or reinstall software
Very clear image
- Very long exposure time.
- Adjustment of your monitor.
(brightness and contrast)
Very dark image
- Insufficient exposure time.
- The tension of the generator is insufficient
(<60kV).
- Adjustment of your monitor.
(brightness and contrast)
Image completely dark
- The active area of the sensor was not exposed to the X-
Rays
- Insufficient X-Ray dose
Image with artifacts or lines
Calibration file doesn't belong to the sensor
- Contact the technical support
Image failure
USB cable failure
- Verify if the USB cable shows any damage indicative
- Verify USB connection with the computer
- Verify if the USB port of the computer has the capacity to
supply enough power for the proper operation of the
sensor.
Fail in the sensor due to physical damage (impact)
- Contact the technical service
In the case of pending issues, contact Alliage authorized technical service.
CABLE REPLACEMENT
Before starting the cable replacement procedure, disconnect the sensor
from the USB port of the computer to avoid damage to your equipment.