The compliance with EMC and EMI regulations can't be guaranteed by
the use of modified cables or those which doesn't comply with the
same standards under what the equipment was validated.
This equipment must not be used adjacent or supported by other
equipment. The recommendations of this manual must be followed.
Don't use accessories, transducers, internal parts of components and
other cables other than those previously specified by the manufacturer.
This may result in increased emission or decreased electromagnetic
immunity and result in improper operation.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm to any part of the Intraoral Dental Digital Imaging Sensor,
including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
To maintain basic safety in relation to electromagnetic disturbances
during the expected service life, always use the equipment in the
specified electromagnetic environment and follow the maintenance
recommendations described in this manual.